Absorbance 96 Mini-Reader

The Absorbance 96 was designed with a focus on providing users with complete flexibility. The outcome is a one-of-a-kind, solid-state microplate reader featuring 96 detection units.

More Info

Absorbance 96 Mini-Reader

The Smallest Plate Reader with Big Performance

The Absorbance Mini-Reader is the most compact spectrophotometer on the market with a footprint only slightly larger than a microplate. It delivers accurate and precise optical density (OD) outputs. The reader is compatible with Automation Trainer’s Validation Reference Plate (VRP) which can be used to test the photometric accuracy and linearity of virtually any reader and validate GLH QC Kit results to achieve NIST-Traceability.

14.5 cm / 5.7 in

Length

9 cm / 3.5 in

Width

5 cm / 2 in

Height

900 g / 2 lbs

Weight

Intuitive Software with Streamlined Navigation

The software boasts a sleek, user-friendly interface with seamless navigation and a logical layout. It avoids hidden windows and cluttered dropdown menus to minimize complexity. Its intuitive design ensures that users can grasp its functions within minutes, eliminating the need for extensive training.

GLH QC Kit Workflow

Prep Plate

Dispense the reference plate and test plate with GLH QC Reagents and mix with plate shaker.

Read Plate

Simply scan the reference plate and the test plate, whether using our GLH QC Absorbance Mini Reader or your own plate reader.

Analyze and Report

GLHTracker software streamlines the entire workflow providing quality-compliant pass/fail results and detailed data for troubleshooting and optimization.

Detection method    
Absorbance   
   
Detection mode   
   
Endpoint, kinetic   
   
Microplates types   
   
96-well microplates   
   
Light source   
   
Up   to 6 LEDs   
   
Detector   
   
96   Photodiodes   
   
Wavelength selection   
   
Example combinations (FWHM 10   nm): 450, 492, 562, 620 nm/405, 450, 540, 630 nm/492, 562, 605, 650 nm. Other   filter combinations in the range of 400–1000 nm possible.    
   
Photometric range   
   
0.0–4.0   OD   
   
Linearity   
   
405 nm: ≤ 1.5 % from 0.0–2.0 OD; ≤ 3.0   % from 2.0–3.0 OD
   
≥ 450 nm: ≤ 1.0 % from 0.0–2.0 OD; ≤   1.5 % from 2.0–3.0 OD    
   
Accuracy   
   
405 nm: ≤ 1.5 % + 0.010 OD from 0.0–2.0   OD; ≤ 3.0 % + 0.010 OD from 2.0–3.0 OD
   
≥ 450 nm: ≤ 1.0 % + 0.010 OD from   0.0–2.0 OD; ≤ 1.5 % + 0.010 OD from 2.0–3.0 OD   
   
Reproducibility   
   
≤ 0.5 % + 0.005 OD from 0.0–2.0 OD; ≤ 1 % + 0.010 OD from 2.0–3.0 OD   
   
Resolution   
   
0.001 OD   
   
Read time   
   
Down   to 3 sec. at single wavelength   
   
Data output   
   
USB   2.0 interface with PC.   
   
Dimensions reader   
   
9.6 x 15.4 x 5.5 cm (W x L x H)   
   
Weight reader   
   
900   g   
   
Power supply   
   
USB   connection 5 V   
   
Power consumption   
   
2.5   Watts   
   
Warranty   
   
The   GLH QC Kit is for laboratory research use by trained users. Determination of   their suitability for specific end-use is the responsibility of the user, who   assumes all liability for loss or damage arising out of the use of the   product. Warranty is limited to   replacement of defective materials, if returned with authorization, within   1-year of purchase date.   
Get in touch

Looking to enhance your automated liquid handling process?